9. Fieldbus interface | 9.2 - Scale protocol | 115
9.2.3.2 Writing control bits
Some device functions can be executed by setting bits directly in bytes 6 and
7 (control bytes) of the write window.
Action of the master Slave reaction
Set bits in the control byte.
Function is executed.
Reset bits in the control byte.
9.2.4 Waiting for the result of the action
When an action requiring more time is started, the end of execution can also
be waited for.
Action of the master Slave reaction
For setting bits, see ChapterWrit
ing bits or Writing control bits.
Set the CmdBusy bit.
Function is executed.
In the event of an error: Set the
CmdError bit and the LastError
byte.
If the function is executed or
timeout:
reset the CmdBusy bit.
Wait until CmdBusy = 0.
Check the CmdError bit.
If CmdError is set:
Evaluate the LastError (for func
tion number4, see ChapterFunc
tion number 4: adjustment infor
mation, error byte (read))