11.1 Error messages measuring circuit
The internal weighing electronics can generate error messages, which are
shown on the display.
Display Error Possible cause
ERR 2 The measuring signal is higher than
Max + (x d) (Overload)
Wrong setting.
Too much weight on the scale.
ERR 3 Measuring input open
The measuring signal is higher than
the permissible range of 36mV.
Measuring cable is interrupted (cable
break detection).
Other hardware defect.
ERR 6 No sense voltage
Load cells not connected.
Sense line or supply line is interrupted.
Wrong polarity or sense voltage is low.
ERR 7 Negative measuring signal
Wrong polarity of load cell signal.
Wrong polarity of load cell supply volt
age.
ERR 9 The measuring signal is higher than
the permissible range of 36mV. Can
not read weight values from ADC
(analog-digital converter).
Internal weighing point:
Error in weighing electronics board.
Defective load cell.
Cable break.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Minebea Intec MiNexx-link-E PR 5215/01 and is the answer not in the manual?

Minebea Intec MiNexx-link-E PR 5215/01 Specifications

General IconGeneral
BrandMinebea Intec
ModelMiNexx-link-E PR 5215/01
CategoryIndustrial Electrical
LanguageEnglish

Related product manuals