6. Operating | 6.3 - WEB menu | 79
Adjustment ·
– New
– Max. capacity
Input: Weight value with decimal place, mass unit
– Scale interval
Selection: corresponding value
– Dead load
Selection: By load, By mV/V
– Adjusted at
Selection: By load, By mV/V, By LC data
– Modify
– Dead load
Selection: By load, By mV/V
– Adjusted at
Selection: By load, By mV/V, By LC data
– By LC data
– Number of loadcells
Selection: 1…8
– Max. capacity of loadcell
Input: Max. capacity E
max
of one load cell
– Gravity
Input: Gravity at the place of installation; default is the value for
Hamburg, Germany: 9.81379 m/s
2
– Hysteresis error
Selection: not specified, specified
– Correction A
Only by selection "specified"
Input: from the calibration certificate of the load cell under
"Hysteresis Correction A"
– Correction B
Only by selection "specified"
Input: from the calibration certificate of the load cell under
"Hysteresis Correction B"
– Certified data
Selection: All load cells same, Each load cell specific
– LC output at max. capacity
Only by selection "All load cells same"
Input: from the calibration certificate of the load cell under "Output at
max. capacity"