6. Operating | 6.3 - WEB menu | 79
 Adjustment   · 
 New
 Max. capacity
Input: Weight value with decimal place, mass unit
 Scale interval
Selection: corresponding value
 Dead load
Selection: By load, By mV/V
 Adjusted at
Selection: By load, By mV/V, By LC data
 Modify
 Dead load
Selection: By load, By mV/V
 Adjusted at
Selection: By load, By mV/V, By LC data
 By LC data
 Number of loadcells
Selection: 1…8
 Max. capacity of loadcell
Input: Max. capacity E
max
of one load cell
 Gravity
Input: Gravity at the place of installation; default is the value for
Hamburg, Germany: 9.81379 m/s
2
 Hysteresis error
Selection: not specified, specified
 Correction A
Only by selection "specified"
Input: from the calibration certificate of the load cell under
"Hysteresis Correction A"
 Correction B
Only by selection "specified"
Input: from the calibration certificate of the load cell under
"Hysteresis Correction B"
 Certified data
Selection: All load cells same, Each load cell specific
 LC output at max. capacity
Only by selection "All load cells same"
Input: from the calibration certificate of the load cell under "Output at
max. capacity"

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Minebea Intec MiNexx-link-E PR 5215/01 and is the answer not in the manual?

Minebea Intec MiNexx-link-E PR 5215/01 Specifications

General IconGeneral
BrandMinebea Intec
ModelMiNexx-link-E PR 5215/01
CategoryIndustrial Electrical
LanguageEnglish

Related product manuals