EasyManuals Logo

Minebea Intec MiNexx-link-E PR 5215/01 User Manual

Default Icon
150 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #97 background imageLoading...
Page #97 background image
6. Operating | 6.13 - Metrology parameters | 97
[Standstill range]
As long as the weight fluctuations remain within this range, the device is
determined to be stable.
Input: 0.01 to 10.00d
[Tare condition]
Selection: Tare without standstill, Tare after standstill
[Tare timeout]
Timeout for a tare/zeroset command that cannot be executed (e.g. due to
mechanical instability of the scale, incorrect filter setting, resolution too
high, standstill condition too strict).
Input: 0.00 to 25.00s.
At 0.0s taring is only carried out when the scale is already stable.
[Zeroset range]
Define a ±range around the zero point determined by the dead load dur
ing adjustment; within this range
the displayed gross weight can be set to zero by a corresponding ex
ternal command, and
automatic zero tracking is active.
Input: 0.00 to 10000.00d
[Zerotrack]
The zero display is automatically maintained within set limits.
Selection: No, Yes
When [No] is selected, the next three parameters are not shown.
When [Yes] is selected, values for the next three parameters must be en
tered.
[Zerotrack indic. range]
Indication range within which automatic zerotrack compensates for devia
tions.
Input: 0.25 to 10000.00d
[Zerotrack step]
If a weight change exceeds the adjusted value, automatic tracking does
not function any more.
Input: 0.25 to 10000.00d

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Minebea Intec MiNexx-link-E PR 5215/01 and is the answer not in the manual?

Minebea Intec MiNexx-link-E PR 5215/01 Specifications

General IconGeneral
BrandMinebea Intec
ModelMiNexx-link-E PR 5215/01
CategoryIndustrial Electrical
LanguageEnglish

Related product manuals