The hold mode
is not activated
even after the
hold switch is
pressed.
1) Is the stylus indented over the measuring
range?
1) Retract the stylus using up/down
fine feed knob.
The hold mode
is not activated
even after the
mid hold switch
is pressed.
1) Is the stylus position moved to the + side
from the mid hold position?
1) Retract the stylus using up/down
fine feed knob.
Abnormal
sound and
vibration are
generated.
1) Are the supporting legs of the installation
base tightened?
1) Securely install the installation
base.
2) Is the leveling of the installation base
appropriate?
2) Level the installation base.
3) Are the fixing parts of the measuring main
unit with the stand or jig tightened
securely?
3) Securely tighten them.
"S0010 USB
dongle cannot
be recognized"
is displayed.
1) Is the USB hardware key connected to the
PC?
1) Insert the USB hardware key to the
PC, and start up analysis software.
During
operation, the
power has
been
interrupted.
1) Is the power cord disconnected?
1) Connect the power cord correctly.
2) Is not power supply voltage falling to less
than 90V ? (voltage dips, short
interruptions or voltage variations of power
supply are included)
2) An electric supply source is
investigated.
When other equipment was
supplied from the same power
outlet, change other equipment to
another power outlet.