3. Preparation for measurement
No. 99MBB793A
3-15
3.9.1 Calibration procedure (Z-axis gain adjustment)
The operation for calibrations is performed with the analysis software,
"FORMTRACEPAK". Refer to the manual of the analysis software "FORMTRACEPAK
(Measurement)" for further information.
1. Prepare gage blocks.
The following explains an example where a 50 mm gage block and a Cross-travel
table are used.
Place a Cross-travel table on the base.
Set the gage block.
2. Move the stylus onto the top surface of the gage block.
3. Perform a calibration using analysis software. Select [Calibrate...] from the
[Measurement] menu.
The dialog box [Z Gain Adjustment] will be displayed.
50mm (standard
accessory)
10mm (standard
accessory)
[Calibrate...] button

Table of Contents