No. 99MBB793A
3-16
4. Select [Static Calibration], and click the [Measure] button.
The dialog box [Calibration Gage] will be displayed.
5. Enter 50mm in the [Step val] field.
If a data for calibration is already entered, click the [OK] button.
In the entry display, click [C], [5], [0], and then [OK] button.
The dialog box [Calibration Gage Leveling] will be displayed.
6. Level the top surface of the gage.
Click the [Leveling] button.
TIP
No change in settings is necessary, but perform the setting change by clicking the
[Set] button as necessary. Refer to "FORMTRACEPAK (Measurement)" for further
information.
The dialog box [Leveling] will be displayed.
[Measure] button
Click.
[Leveling] button