83BSupported Service Summary 11BAppendix C: Command Summary
110
The method to issue the Coarse_Zero_Offset service request is shown below.
Table 63: Method to Issue the Coarse_Zero_Offset Service Request
Service Code Class Instance Service Data
0x32 0x31 0x01 None, a target of zero will be assumed.
This service is valid only when the Raw Pressure Value is within –4 to +20% of the full scale
range of the device. Refer to S-Analog Sensor Object, page 65, for more information.