Measurement accessory kits
The following measurement accessory kits can be used for a Creep test:
Stand, basic (with any test head and needle or tube adaptor, with or without filter)
Fixture f. ASTM restraining (with any test head and needle or tube adaptor, with or without filter)
Tube adapter
•IV-Bag adapter
Valve Test Unit (VTU)
Handheld needle, mono
Handheld, needle, twin
Creep test settings
For a Lippke 4500 the test parameters are set in the PC program and for a Lippke 4000 the parameters
are set on the device display.
* = parameter only available in the Lippke 4500 PC Software
Names in parenthesises are the corresponding Lippke 4000 parameter names
Parameters:
Date last modification * Filled out automatically
User * Filled out automatically
Name
(Name)
Test name
Name 1 * Can be used to give additional information about the test
Name 2 * Can be used to give additional information about the test
Package type
(Package Type)
- Normal This is the recommended standard setting. Should
be used whenever possible.
- Porous Use for porous packages like Tyvek.
- Sensitive Do not use for a Creep test.
Test pressure [mbar]
(Test pressure)
Pressure at which the test is performed
(Typically 80% of the burst pressure)
Creep time [sec]
(Creep Time)
The period of time where test pressure is maintained.
With oxygen measurement * Requires a PacCheck 3XX - If checked the O
2
level is measured.
Limit oxygen [%]
(O2 Limit)
Upper limit for acceptable O
2
measurement.
With CO2 measurement * Requires a PacCheck 3XX - If checked the CO
2
level is measured.
Limit CO2 [%] * Upper limit for acceptable CO
2
measurement.