Lippke 4000/4500 EN User Guide 51
5. Setting Up
Compressed air connections
Compressed air supply requirements
The compressed air supply must be dry, clean and free from oil - see specifications in "Air supply" on
page 71.
CAUTION! If the specified maximum pressure is exceeded, there is a danger of bursting of
tubing and other components.
NOTE! Operation to within specification is not guaranteed if the Air Supply pressure is
below the specified minimum input pressure.
Filter recommendations
In cases where a filter is required to clean the air enough to fulfil the required specifications, we
recommend the following two types which can be used in most cases:
Festo LFMB-D-MINI (162626)
Zander Min-DRY MDK and KT 2016
NOTE! Even when using the above filter types it is still the responsibility of the owner to
ensure that the compressed air supply fulfils the requirements as specified in "Air
supply" on page 71.
Correct mounting/dismounting of fittings
CAUTION! When mounting/dismounting fittings in order to connect compressed air hoses
or to replace fittings, it is very important that you hold against on the already mounted
fittings to avoid damaging the existing assemblies.