P/N 330654-E 01/2018 COPYRIGHT © Dansensor A/S
72 User Guide EN Lippke 4000/4500
Accessories data
Stand, basic Dimensions (H x W x D): 407 x 250 x 330 mm (16 x 9.8 x 13”)
Weight: 3.6 kg (7.9 lbs)
Fixture f. ASTM restraining Dimensions (H x W x D): 412 x 400 x 405 mm (16.2 x 15.7 x 15.9”)
Weight: 13.5 kg (29.8 lbs)
Restraining height: 6. 35 - 76.2 mm (¼ - 3”)
in steps of 6.35 mm (¼”)
Max. package size (W x D): 230 x 400 mm (9 x 15.7”)
PPC 300 II Dimensions (H x W x D): 200 x 470 x 495 mm (7.9 x 18.5 x 19.5”)
w. Package fixture Weight: 20.0 kg (44.1 lbs)
Restraining height: 31 - 80 mm (1.2 - 3.1”)
Max. package size (W x D): 310 x 360 mm (12.2 x 14.2”)
Valve Test Unit (VTU) Dimensions (H x W x D): 115 x 176 x 189 mm (4.5 x 6.9 x 7.4”)
Weight: 2.6 kg (5.7 lbs)
Max. sample size Ø 24.1 mm (0.95”)
Basic specifications
Ambient operation pressure 900 - 1050 mbar (13 - 15.2 psi)
Measuring ranges 1: 10 - 1000 mbar (0.5 - 14.5 psi)
2: 200 - 3000 mbar (2.9 - 43.5 psi)
Resolution per 1: 0.1 mbar (0.0015 psi)
Measuring range 2: 1.0 mbar (0.015 psi)
Accuracy per 1: ± 0.5 mbar (0.007 psi) or 1% *
Measuring range 2: ± 2.0 mbar (0.030 psi) or 1% *
* whichever is greater and within specified range.
Can be used below with reduced specifications.
Reproducibility per 1: ± 0.5 mbar (0.007 psi) or 1% *
Measuring range 2: ± 2.0 mbar (0.030 psi) or 1% *
* whichever is greater and within specified range.
Can be used below with reduced specifications.
Ambient temperature Operational: +15 to +45 °C, less than 90% RH, non-condensing
Storage: -10°C to +60°C, less than 95% RH, non-condensing
Pressure units mbar, mmHG or psi
Test time 1 - 9.999 sec.
Interface languages English, German, French, Spanish

Table of Contents