Lippke 4000/4500 EN User Guide 49
Measurement accessory kits
The following measurement accessory kits can be used for a Bubble test:
Stand, basic (with any test head and needle or tube adaptor, with or without filter)
Tube adapter
•IV-Bag adapter
Handheld, needle, twin
Bubble test settings
The test parameters are set in the PC program.
Parameters:
Date last modification Filled out automatically
User Filled out automatically
Name Test name
Name 1 Can be used to give additional information about the test
Name 2 Can be used to give additional information about the test
Package type - Normal This is the recommended standard setting. Should
be used whenever possible.
- Porous Use for porous packages like Tyvek.
- Sensitive Not used for Bubble tests
Test pressure [mbar] Pressure at which the test is performed
Test time [sec] The period of time where test pressure is maintained.
Displayed limits in chart:
Pressure, upper limit in chart
[mbar]
Upper pressure value for the [mbar] axis in the resulting graph chart
Pressure, lower limit in chart
[mbar]
Lower pressure value for the [mbar] axis in the resulting graph chart