34 | P a g e
CAUTION Variable recirculating blowers are loud and operate between 65-82 decibels (dBa). Long-term noise
exposure requires the use of hearing protection.
A notification will pop up if the recirculation fan speed is increased past 75% of full speed warning that blower noise
over 80 decibels requires hearing protection for long-term exposure. At higher fan speed the home screen will indicate a
red icon visually alerting the indicating hearing protection is required. The icon can be clicked on for further information
of the icon and warning.
Alarm notification must be setup to utilize the full functionality of the alarming features. Reference section
6.1.9 below for instructions and descriptions of the different alarm types and how to set up.
6.1.5 Ventilation Button
When the ventilation button is pressed, the user is taken to the ventilation screen. This screen features the Recirculation
fan and Front Air System (FAS) On and Off toggles. The patented FAS system is a key safety feature on Mopec laboratory
equipment. The principle of operation is simple and safe. The FAS uses a series of fans to provide supplemental airflow
at the front edge of the workstation directed at the backdraft vent. This helps push and pull any outgassed fumes from
the worksurface and safely away from the user.