The Integrated Filtration System page is intended to set a reminder and alert the user(s) when it is time to replace
the filters in the system. This functionality can be used for units with standard particulate filters and Potassium
Permanganate filters that are utilized with grossing stations having recirculation. To set a determined Filter Change
Interval tap on the HOURS box, input the required hours, and select Submit. This will save the interval. To reset the
time remaining simply press RESET TIMER. It will not display the time remaining from your set interval.
Units with in-house ventilation you need to be aware that the filter change frequency will be greater as in-house
typically runs 24/7.