REF 9515-151-50-ENG Rev S1
X-Scribe
STRESS EXERCISE TESTING SYSTEM
USER MANUAL
Manufactured by Mortara Instrument, Inc., Milwaukee, Wisconsin U.S.A.
CAUTION: Federal law restricts this device to sale by or on the order of a physician.

Table of Contents

Questions and Answers

  • J
    Julie FriedmanSep 7, 2025
    What does 'Unknown Device' mean on Mortara Medical Equipment?
    • V
      vincentmitchellSep 7, 2025
      If the system displays “Unknown Device” or “Unknown Manufacturer”, the TRPC or EPC is not recognized due to an incorrect installation or lack of installation of the device driver. Note that TRPC or EPC is not compatible with X-Scribe V3.10. You will need an MMM device and drivers.
  • C
    Catherine MartinezAug 22, 2025
    How to fix treadmill belt slippage on Mortara Medical Equipment?
    • S
      stephanie70Aug 22, 2025
      If the treadmill belt starts to slip, it could be due to looseness that allows it to shift. To resolve this, tighten the adjustment bolts on both sides until the slippage stops.
  • L
    Latoya HernandezSep 5, 2025
    What to do if there is no signal onscreen on Mortara X-Scribe Medical Equipment?
    • S
      stacyharmonSep 5, 2025
      If there is no signal onscreen (wireless option) with the patient connected (or with a simulator), and only the squared waveform/lead fail indicator appears, check that X-Scribe and the digital transmitter work on the same channel. Refer to the transmitter user manual to check and change the channel. Refer to Settings in this manual to set the receiving channel for X-Scribe.
  • J
    Julie RuizSep 3, 2025
    How to troubleshoot packet errors on Mortara Medical Equipment?
    • C
      Charles CampbellSep 3, 2025
      If you are experiencing packet errors or signal drop (wireless option), and the signal onscreen shows artifact even when tested with a simulator, check that no medical equipment is operating at the same frequency in proximity to the system. Run the built-in spectrum analyzer to locate the source of the interference and to set the proper channel on the digital transmitter.
  • M
    Mark RoseSep 2, 2025
    What to do if Mortara X-Scribe Medical Equipment menu cursor won't move?
    • D
      Dana RodriguezSep 2, 2025
      If the menu cursor will not move, the mouse cable may be disconnected or the keyboard/mouse cables may be transposed. Turn the system OFF and check the mouse port connections.
  • B
    bbrownSep 1, 2025
    Why is my Mortara X-Scribe keyboard not responding?
    • J
      James LopezSep 1, 2025
      If there is no response to keyboard commands, the keyboard cable may be disconnected or the keyboard/mouse cables may be transposed. Turn the system OFF and check the keyboard to mouse port connections.
  • L
    Lauren TorresAug 30, 2025
    Why is my Mortara Medical Equipment Z-fold printer not working?
    • H
      Heather AguilarAug 30, 2025
      If the Z-fold writer paper is out, the light on the Z-fold writer is on, and it's not printing or printing ECGs/reports unevenly, it could be due to a paper jam, no paper in the tray, an open writer door, or a printer head that needs cleaning. Open the writer cover and remove any jammed paper. Insert a new packet of paper in the tray. Make sure the writer door is latched. Refer to printer head cleaning instructions in Appendix F.
  • M
    Michael PhamAug 29, 2025
    What to do for ECG Front-End Error on Mortara X-Scribe?
    • L
      Lisa DavisAug 29, 2025
      If the ECG Front-End Error occurs, and you can’t proceed, it could be that the MMM device is not correctly connected or that there are incorrect Power Savings settings. Try disconnecting the USB cable from the PC and then reconnecting it. The system should emit a confirmation beep. If rebooting fixes the problem, the system likely has incorrect Power Saving settings. To correct this: 1. Close the X-Scribe application. 2. Select “Start”, “Settings”, then “Control Panel”. 3. Double click on “Power Management” icon. 4. If displayed, set “Stand-by” setting to “Never”. 5. Click OK and close all open Windows. 6. Reboot system.
  • L
    Lori RobertsAug 27, 2025
    How to correct baseline drift on Mortara Medical Equipment?
    • L
      lpughAug 28, 2025
      Baseline drift can be caused by poor skin-to-electrode contact. Re-prep the skin and replace the faulty electrode to resolve this issue.
  • M
    Misty BlackAug 27, 2025
    How to reduce muscle noise in Mortara Medical Equipment readings?
    • A
      Ashley HaneyAug 27, 2025
      Muscle noise can occur if the electrode is placed over muscle or fatty tissue. To fix this, find a stable electrode site, re-prep the skin, and apply a new electrode.

Summary

User Safety Information

Warning and Caution Symbol Definitions

Defines symbols for warnings, cautions, and notes to ensure safe operation.

General Safety Precautions

Provides essential safety guidelines for operating the device.

Electrical and Patient Safety Guidelines

Ensures electrical safety and patient protection during use.

Specific Operational Hazards and Warnings

Highlights hazards like defibrillation, flammable anesthetics, and improper use.

Introduction Section 1

Manual Purpose and Audience

Outlines the manual's scope and intended reader.

Intended Use and Indications for Use

Defines the device's purpose and when it should be used.

System Description and Components

Details the X-Scribe system, its features, and included parts.

X-Scribe Specifications Overview

Lists the technical specifications of the X-Scribe system.

Setup and Installation Section 2

X-Scribe System Setup and Installation

Step-by-step guide for assembling and installing the X-Scribe system.

Interconnect Diagram and Isolation Transformer

Visual guide to system connections and details on the isolation transformer.

Using the X-Scribe Section 3

Starting the X-Scribe Application

Instructions for powering on and launching the X-Scribe software.

Key Features and Capabilities

Overview of the X-Scribe's functions and features.

User Interface and System Display

Explanation of the graphical user interface and on-screen data.

Function Key Operations and Controls

Details the use of function keys for system control.

Conducting a Stress Test Section 4

Patient Preparation and Hookup Procedures

Steps for preparing the patient and attaching electrodes.

Using Filters for Signal Quality

Guides on applying filters like SCF, AC, and 40 Hz.

Conducting the Test: Phases and Workflow

Details on test phases: Pre-Exercise, Exercise, Recovery, and Final Report.

Report Management and Archiving

Procedures for managing, printing, and archiving test reports.

System Configuration Section 5

System Settings and RF Scan

Configuring AC frequency, RF channels, and performing RF scans.

User Profile and Exercise Protocol Management

Creating, modifying, and selecting user profiles and exercise protocols.

Network Configuration and User Setup

Setting up network data export/import and standard user configurations.

Maintenance and Troubleshooting Appendix A

Routine Maintenance and Cleaning Instructions

Guidelines for regular equipment cleaning and upkeep.

Troubleshooting Chart and System Information Log

Diagnosing issues and recording system details for service.