MST Module G Status B (revised Version):
Connection of 2 racks:
Fig. 99
The CAN terminating resistor at Rack 1 is not set, at Rack 2 it is set.
Connection of 3 racks:
Fig. 100
The CAN terminating resistor at Rack 1 and Rack 2 is not set, at Rack 3 it is set.
Connection of 4 racks:
Fig. 101
The CAN terminating resistor at Rack 1, Rack 2 and Rack 3 is not set, at Rack 4 it is set.
For every further rack, a T-piece, ribbon cable and a CAN line socket/plug is needed.