US
6.3 Replacement of Sensors
Sensors which are no longer able to reach the minimum signals, the zero point of which can no
longer be adjusted, or which no longer function properly for some other reason must be replaced.
For this purpose, inhibit the corresponding measure point in the “Setup/Measure Points” menu.
Then perform the following steps:
- Remove the connector plug of the sensor from the MAT/MAT TS module or the sensor cable
from the MGT 40 TS module.
- Pull out the old sensor, put the new [properly functioning] sensor in its place, and install it.
- Reconnect the connector plug of the sensor to the MAT/MAT TS module or the sensor cable
to the MGT 40 TS module.
- Perform a first calibration as described in Chapter 6.1 , under consideration of the required
sensor recovery time.
- Check the configuration of the sensors in the “Setup/Measure Points” menu.
- Unlock the measure point.