EasyManua.ls Logo

MSA SUPREMATouch - Page 84

MSA SUPREMATouch
309 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MSA AUER
MSA
Maintenance and Service
SUPREMATouch
83
US
- Set the output current of the power source to 4 mA.
- Unlock the measure point in question in the “Setup/Measure Points” menu to allow a calibra-
tion.
- Select the “Maintain/Calibration” menu.
- Select the measure point to be calibrated in the “Measure Point” field.
- Enter 0 % of the measuring range in the field “Zero Gas” as the zero gas concentration.
- Enter 100 % of the measuring range in the field “Test Gas” as the test gas concentration.
- Begin the calibration with the “Start” button.
- After actuating the “Start” button, you will be asked for the password at the first measure point.
Enter the password or actuate the key switch.
- The “End Calibration” submenu will appear.
- The current measurement value UA of the measure point to be calibrated will appear in the
“Sig:” field. For an input current of 4 mA, a value of 400 mV ± 10 mV should be displayed here.
In the field “Ux=” no value or ***** is displayed.
- If the value UA is within the tolerance range [400 mV ± 10 mV], confirm by actuating the "Store"
button. The value will be appear in the field “CAL-ZERO / mV”.
- Set the output current of the power source to 20 mA.
- The current measurement value UA of the measure point to be calibrated will appear in the
“Sig:” field. For an input current of 20 mA, a value of 2.000 mV ± 10 mV should appear here.
- If the value UA is within the tolerance range [2000 mV ± 10 mV], confirm by actuating the
"Store" button. The value will appear in the field “CAL-SPAN/ mV”.
- End the calibration of the selected measure point by actuating the "End" button
- Set the power source back to 4 mA and lock the measure point again.
- Cut the connection between the MAT module terminal and the power source and reconnect
the sensor.
- After allowing the sensor to recover sufficiently, unlock the measure point.
- In the beginning calibration menu, select the next measure point and repeat the procedure de-
scribed above.

Table of Contents

Related product manuals