OPR360US H0711
5-10
5
Alarms
Alarm or
Device Alert
Message
• Priority
• Type
•
Lamp Color
Description
and/or
Criteria
Remedy
Control CPU
Failed
• High
• Device Alert
• Red
Control CPU
Failure
Check patient and provide
an alternate source of
ventilation.
Contact a qualied service
representative.
Control uP
Failed
• High
• Device Alert
• Red
Monitor
Commun-
ications Failure
The control
processor does
not respond to
a request from
the monitor
processor.
The control
processor is not
running.
Check patient and provide
an alternate source of
ventilation.
Contact a qualied service
representative
Control RAM
Failed
• High
• Device Alert
• Red
Control RAM
Failure
Random access
memory that
is used by
the control
processor on
the main PCB is
damaged.
Check patient and provide
an alternate source of
ventilation.
Contact a qualied service
representative.
Control
Tasks Failed
• High
• Device Alert
• Red
Control Task
Continuity
Failure
Software tasks
of the control
processor have
operated out of
sequence.
Check patient and provide
an alternate source of
ventilation.
Contact a qualied service
representative.
Data Read
Failure
• High
• Ventilator
• Red
Error is
detected during
the reading of
stored data at
power up.
Contact a qualied service
representative.