Alarms
5
Alarm or
Device Alert
Message
• Priority
• Type
Lamp Color
Description
and/or
Criteria
Remedy
Device Alert
(If message
display is
possible)
Unsilence-
able audible
alarm.
Device Alert
indicator
flashes if
possible.
• High
• Device Alert
• Red
Ventilator
malfunction.
Also activated
if less than
10% of the
internal battery
operation time
remains when
ventilator is
operating on
the internal
battery.
Check patient and provide
an alternate source of
ventilation.
If the alarm is due to
battery depletion, plug the
ventilator into an alternate
power source to allow
charging.
If not a result of a low
battery, contact a qualified
service representative.
Dual RAM
Failed
• High
• Device Alert
• Red
Dual RAM
Failure
Random access
memory that
is shared by
the control
and monitor
processors has
failed.
Check patient and provide
an alternate source of
ventilation.
Contact a qualied service
representative.
EEPROM
Read Error
• Low
• Ventilator
• None
This error is
logged into the
Event History
Log when the
software type
read from the
EEPROM on the
main board is
invalid.
No audible or
visual alarm is
displayed, and
no message
appears on the
GUI.
EEPROM Read Error
is a Technical Service
Message that does not
affect functionality of
the ventilator. Contact
a qualified service
representative.
F
IO2 High
• Medium
• Ventilator
• Yellow
Measured FIO2
is more than
0.07 above the
FIO2 setting for
30 seconds.
(cont. next
page)
Verify that the source
gases and connections
are secure and functional.
Verify that the oxygen
source gas is providing
100% oxygen.
(cont. next page)