OPR360US H0711
5-12
5
Alarms
Alarm or
Device Alert
Message
• Priority
• Type
•
Lamp Color
Description
and/or
Criteria
Remedy
FIO2 High
(cont.)
• Medium
• Ventilator
• Yellow
NOTE: This
alarm is
inactive if the
e360 detects
a disabled
or defective
oxygen sensor,
or oxygen
supply pressure
is below 30
psig.
Calibrate the O
2 sensor
by pressing the O2 3 min.
button or by accessing O2
Sensor calibration button
in Setup & Calibration/
Sensors.
Replace O
2 sensor if
necessary.
NOTE: Insufficient supply
inlet pressure of air may
cause ventilator to be
unable to deliver required
high flow rate, resulting in
FIO2 alarm.
F
IO2 Low
• High
• Ventilator
• Red
Measured FIO2
is more than
0.07 below the
FIO2 setting for
more than 30
seconds.
NOTE: This
alarm is
inactive if the
e360 detects a
disabled or
defective
oxygen sensor,
or oxygen
supply pressure
is below 30
psig.
Verify that the source
gas and connections are
secure and functional.
Verify that the oxygen
source gas is providing
100% oxygen.
Verify that ventilator
flow and trigger settings
are adequate to meet
patient’s needs.
Calibrate the O2 sensor
by pressing the O2 3 min
button or by accessing O2
Sensor calibration button
in Setup & Calibration.
Replace O2 sensor if
necessary.
NOTE: Insufficient supply
inlet pressure of O2 may
cause ventilator to be
unable to deliver required
high flow rate, resulting in
FIO2 alarm.