Alarms
5
Alarm or
Device Alert
Message
• Priority
• Type
•
Lamp Color
Description
and/or
Criteria
Remedy
Insp Time
Too Long
(cont.)
• Medium
• Ventilator
• Yellow
in an inspiratory
time greater
than 5 seconds,
including any
pause time, so
ventilation is not
delivered
according to
settings.
and Pause settings and
readjust as necessary.
Insp Time
Too Short
• Medium
• Ventilator
• Yellow
Ventilator
settings or
alarm limit
violation
results in an
inspiratory time
< 0.1 seconds
for adult or <
0.05 seconds
for pediatric
patients,
excluding
any pause or
inspiratory hold.
While in VC,
an Insp. Time
setting that
cannot deliver
the set Tidal
Volume.
Evaluate Tidal Volume and
Flow settings and readjust
as necessary.
Potential cause may be
High Paw alarm violation.
Resolve as needed.
Low
Baseline
Pressure
• High
• Patient/
Ventilator
• Red
Monitored
airway pressure
has been below
the PEEP/CPAP
setting for
more than 0.5
seconds for two
consecutive
breaths.
Turn Leak Compensation
ON.
Check breathing circuit for
leaks or disconnects.
If there is an airway/mask
leak, turn ON Noninvasive
Ventilation..
The alarm is inactive when
PEEP/CPAP is set below
4 cmH
2O/mbar.
Low Battery
• High
• Ventilator
• Red
Internal battery
operating
(cont. next
page)
Connect the ventilator to
any AC power source to
allow the internal battery
to recharge.
(cont. next page)