OPR360US H0711
5-16
5
Alarms
Alarm or
Device Alert
Message
• Priority
• Type
•
Lamp Color
Description
and/or
Criteria
Remedy
Low Battery
(cont.)
• High
• Ventilator
• Red
capacity has
dropped
to 25% of
capacity.
NOTE: The ventilator
continues to operate
normally until the battery is
depleted, at which time a
Device Alert alarm occurs.
WARNING! To ensure
continued ventilator
operation, substitute an
alternate power source,
such as AC power,
immediately when a Low
Battery alarm occurs.
Low MVE
• High
• Patient
• Red
Monitored MVE
< set
Low MVE alarm
limit.
Check patient, evaluate
ventilation and alarm
settings and readjust if
necessary.
Check breathing circuit for
leaks or disconnects.
Check that the exhalation
flow sensor is installed
correctly.
The ventilator provides
back up ventilation while
this alarm limit is violated.
Low Paw
• High
• Patient/
Ventilator
• Red
Monitored
airway pressure
does not
reach the Low
Paw alarm
limit during
a mandatory
inspiration for
two breaths
(does not apply
to manual,
spontaneous
or pressure
supported
breaths).
Check breathing circuit for
leaks or disconnects.
Evaluate ventilation
and alarm settings and
readjust if necessary.
The alarm is recovered
when Ppeak for a
mandatory breath
(including Back Up
Ventilation) is greater than
Low Paw alarm limit.