OPR360US H0711
5-17
Alarms
5
Alarm or
Device Alert
Message
• Priority
• Type
•
Lamp Color
Description
and/or
Criteria
Remedy
Low Paw
Below PEEP
Informational
Message
Set Low Paw
Alarm level is
less than set
PEEP
Change alarm limit to
greater than the set PEEP
level.
M Internal
System
• High
• Device Alert
• Red
Control
Exception
Failure
The Control
processor
has detected
an abnormal
operation
such as illegal
instruction
or divide by
zero that was
generated by
the control
software.
Check patient and provide
an alternate source of
ventilation.
Contact a qualied service
representative.
Monitor uP
Failed
• High
• Device Alert
• Red
Control
Commun-
ications Failure
The monitor
processor does
not respond
to a request
from the control
processor.
The monitor
processor is not
running.
Check patient and provide
an alternate source of
ventilation.
Contact a qualied service
representative.
Monitor CPU
Failed
• High
• Device Alert
• Red
Monitor CPU
Failure
The monitor
processor on
the main PCB is
bad.
Check patient and provide
an alternate source of
ventilation.
Contact a qualied service
representative.
Monitor RAM
Failed
• High
• Device Alert
• Red
Monitor RAM
Failure
(cont. next
page)
Check patient and provide
an alternate source of
ventilation.
Contact a qualied service
representative.