OPR360U A0509
5-1
5
OPR360U A0509
5-1
5
OPR360U A0509
5-1
5
OPR360U A0509
5-1
5
OPR360US H0711
5-18
5
Alarms
Alarm or
Device Alert
Message
• Priority
• Type
Lamp Color
Description
and/or
Criteria
Remedy
Monitor RAM
Failed (cont.)
• High
• Device Alert
• Red
Random
Access Memory
that is used
by the monitor
processor on
the main PCB is
damaged.
Monitor
ROM Failed
• High
• Device Alert
• Red
Monitor ROM
Failure
Read only
memory that
stores the code
of the monitor
processor
has detected
an illegal
checksum.
Check patient and provide
an alternate source of
ventilation.
Contact a qualied service
representative.
Mon Task
Failed
• High
• Device Alert
• Red
Monitor Task
Continuity
Failure
Software tasks
of the monitor
processor have
operated out of
sequence.
Check patient and provide
an alternate source of
ventilation.
Contact a qualied service
representative.
No
Commun-
ication
• High
• Ventilator
• none
Continuous
audible alert.
This message
is recorded
into the Event
History Log
when there is a
communication
failure to the
touch screen.
No visual alarm
displayed, and
no message
appears on the
GUI.
If the communication is
reestablished the alert will
automatically reset.
No Communication
is a Technical Service
Message that does not
affect functionality of
the ventilator. Contact
a qualified service
representative if the alert
does not reset.