EasyManua.ls Logo

Newport e360 - Page 93

Newport e360
150 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
OPR360U A0509
5-1
5
OPR360U A0509
5-1
5
OPR360U A0509
5-1
5
OPR360U A0509
5-1
5
OPR360US H0711
5-19
Alarms
5
Alarm or
Device Alert
Message
• Priority
• Type
Lamp Color
Description
and/or
Criteria
Remedy
O
2 Sensor
Disconnect
Informational
Message
The O2
Sensor is
disconnected.
Reconnect the O2 sensor.
NOTE: Always use an
external O2 monitor with
alarms while ventilating
with this message.
O2 Sensor
Error
• Medium
• Ventilator
• Red
Ventilator
cannot calibrate
the O2 sensor
correctly.
Change the sensor as
soon as possible.
NOTE: Always use an
external O
2 monitor with
alarms while ventilating
with this message.
O
2 Supply
Loss
• High
• Ventilator
• Red
Oxygen supply
inlet pressure is
below minimum
supply
pressure
required.
Check that oxygen supply
is connected and providing
>/= 30 psig at the
ventilator inlet, especially
during inspiration.
Provide alternate
ventilation if necessary.
The ventilator continues
breath delivery using the
remaining air gas supply,
and does not calibrate the
O2 sensor.
Out of
Range
Informational
Message
A setting
adjustment is
out of range
for the patient
category
selected.
Re-adjust the setting to
within range or change the
patient category.
Power
Failure
• High
• Device Alert
• Red
Power Failure
DC power out
of tolerance.
Check +12
VDC, -12 VDC,
and +5 VDC.
(cont. next
page)
Check patient and provide
an alternate source of
ventilation.
Contact a qualied service
representative.

Table of Contents

Other manuals for Newport e360

Related product manuals