Alarms
Alarm or
Device Alert
Message
• Priority
• Type
Lamp Color
Description
and/or
Criteria
Remedy
Power
Failure
(cont.)
• High
• Device Alert
• Red
(The e360 may
have been
powered by the
internal battery
until it was
depleted and
a Device Alert
resulted.)
Pressure
Limit Below
PEEP
Informational
Message
The set
pressure limit
is less than the
set PEEP.
Adjust either set Pressure
Limit or set PEEP so
that the Pressure Limit is
above the PEEP level.
Pressure
Support +
PEEP > 60
Informational
Message
The sum of
Pressure
Support plus
the PEEP
Settings is
greater than 60
cmH
2O/mbar.
Adjust either Pressure
Support or PEEP level to
less than a combined total
of 60 cmH2O/mbar.
Shutdown
Alarm
No Indicator
No message
is associated
with this alarm.
The Shutdown
alarm provides
confirmation that
the ventilator
power state has
been switched
to OFF.
Press the Alarm Silence
button to cancel the
Shutdown alarm.
Sustained
High
Baseline
Pressure
• High
• Ventilator
• Red
Monitored
PEEP pressure
has been >/=
8 cmH
2O/
mbar above set
PEEP/CPAP for
over 6 seconds
in Ped/Infant or
10 seconds for
Adult.
(cont. next
page)
Check circuit tubing for
occlusions and/or fluids.
Check expiratory lter
for increased resistance/
blockage, replace if
necessary.
Evaluate ventilator settings
and readjust if necessary.
Check to verify that the
Exhalation Valve is clean/
free of debris. Replace if
necessary.
If not resolved, use
alternate means of
ventilation.