Alarms
5
Alarm or
Device Alert
Message
• Priority
• Type
Lamp Color
Description
and/or
Criteria
Remedy
Sustained
High
Baseline
Pressure
(cont.)
• High
• Ventilator
• Red
WARNING!
Ventilation and
triggering are
suspended,
the exhalation
valve and the
emergency
relief valve open
as needed for
a minimum of
5 seconds to
relieve pressure
and no flow is
delivered until
the condition is
resolved.
Ventilation
Suspended
Informational
Message
The Suction
Disconnect
feature is
enabled.
Informational only.
Volume
Target Not
Met
• Medium
• Patient/
Ventilator
• Yellow
In Volume
Target
Pressure
Control or
Volume Target
Pressure
Support breath
type, the set
tidal volume
cannot be
delivered within
the set Pressure
Limit for two
consecutive
breaths.
Check patient condition
for reversible cause.
Increase t Insp, reduce
Resp Rate, increase
Slope/Rise, adjust Exp.
Threshold, increase
Pressure Limit or reduce
Tidal Volume if necessary.
Factors that could trigger
the alarm: agitation, biting
of ET tube, coughing,
increase in expiratory lter
or patient resistance or
drop in compliance (e.g.
secretions,
pneumothorax).