Characteristics
3. Characteristics
3.1 Description of the sterilizer
The DXPclave is a table-top steam sterilizer designed for the sterilization of dental and medical products and
equipment, in accordance with the requirements of standard EN 13060.
It consists of an airtight stainless steel sterilization chamber accessed through a front door; it is protected by an
external shock-resistant moulded plastic body and equipped with protective devices that allow operators to use
it in full safety. The sterilization cycles are started from the operator touch-screen controlpanel on the front of the
device, beside the door.
A detailed description of the units that make up the sterilizer and the components supplied is given in the following
paragraphs.
3.2 Intended use
Small steam sterilizer used to sterilize medical devices or for materials that are likely to come into contact with
blood or body fluids, according to requirements of EN 13060:2014+A1:2018 it is suitable for following sterilization
cycle types and loads:
Class B sterilization
Sterilization of solid products, porous products and lumen devices, wrapped or non-wrapped as represented by
the test loads.
Class S sterilization
Sterilization of non-wrapped solid products.
Sterilizing instruments unsuitable for this process may expose the operator to risk, cause damage to the
sterilizer and compromise its safety devices.
Always check the manufacturer’s label to ensure that products are suitable for sterilization.
The device is not suitable for the sterilization of liquids and flammable materials.
Do not use the device in the presence of anaesthetic or flammable gases.
The room where the device is installed should be adequately ventilated to prevent excessive humidity.
Any possible condensation is collected in the small drip tray.
3.3 Environmental conditions
The sterilizer is designed to operate in medical locations (e.g. Dental practices, dental and medical
clinics, hospitals) with following environmental conditions:
temperatures between 5°C and 40°C;
maximum relative humidity of 80% with environmental temperature until 32°C and linearly decrease until 50%
in case of environmental temperature until 40°C;
air pressure between 750 mBar and 1050 mBar;
altitude between 0 and +2000 meters above sea level.
Transport and storage conditions: temperature -10°C÷50°C, humidity without condensation 10÷95%, air pressure
50÷106 kPa.