49
Alarms
i23 Check clean water quality
Conductivity of clean water
higher than 15 µS/cm, therefore
too high.
Replace or fill the clean water
tank with demineralized water.
i30 Wait for cooling
Chamber temperature too high,
Vacuum test not possible.
Turn off the device and leave the
door open to cool the chamber.
i50 Cycle interrupted
The cycle was manually
interrupted by the operatore.
Sterlization is not completed.
Dry the chamber, if wet, and
restart the cycle. If it persists,
contact technical assistance.
i60 Drying interrupted
The cycle was manually
interrupted by the operatore
during the drying phase.
Dry the load or use it
immediately.
6.3 Notification list
NR. DISPLAY MESSAGE CAUSE SOLUTION
N04
Replace the bacteriological
filter
Clogged bacteriological filter.
Replace the bacteriological filter
and reset the dedicated counter.
N05 Clean the drain filter Clogged drain filter.
Clean or replace the filter under
the metal plate at the bottom of
the chamber.
N31 Clean the door gasket
Vacuum test failed. Possible
leak in the door gasket.
Clean the door gasket and
repeat the test. If it persists,
contact technical assistance.
N32 Vacuum test failed: try again
The expected vacuum value was
not reached within the set time.
Dry the chamber, if wet, and
restart the cycle. If it persists,
contact technical assistance.
N33 Clean the door gasket
Dual Helix test failed. Possible
leak in the door gasket. The
pressure has not decreased in
the expected time.
Clean the door gasket and
repeat the test. Pressure loss.
If it persists, contact technical
assistance.
N40 Pre-heating phase too long Slow heating of the room.
Run the cycle with less load. If
the problem persists, contact
the techincal assistance. Check
the mains voltage.
N61 Reduce load
The minimum pressure value
has not been reached within the
set time.
Reduce the load to make drying
easier.
N70 Faulty conductivity meter
Faulty or defective conductivity
meter.
Contact technical assistance.