Installation
4.3 Initial start-up
The following operations must be carried out by qualified and properly trained personnel.
Incorrect procedures and settings can jeopardize the quality of sterilization and cause hazards.
Check that the power supply has the right voltage and plug the power cord
into the outlet.
Switch on the device by the main switch.
Fill the demineralized water tank up to the maximum level. When the tank is
full, the device signals it with an acoustic signal.
The use of poor quality demineralized water can leave limestone on the instruments, on the inside of the chamber
and on the trays. It is recommended to use demineralised water with a conductivity <15 µS/cm in accordance with
the EN 13060 standard. Do not use domestic tap water, even if treated with a filter or water softener.
Do not use battery water or other fluids or additives, as these can cause irreversible damage to the device and
hazards to the operator.
For daily inactive periods, the main switch can be left in the ON position.
The door remains locked when the device is switched off; if it is still locked when switched on, Press the unlock
button in the main menu.