38
Using the sterilizer
5.7 Program selection
Depending on the degree of sterilization to be obtained, the operator has various types of programs available,
with a class B or class S sterilization cycle, described in the program table.
To select the programs, press the corresponding button on the operator panel.
PROGRAM TABLE
Program
Parameters
Cycle class
Process
value
Cycle total time
(with maximum load)
Type of Load
Maximum
load*
Sterilization
time
Pre-vacuum
phases
Drying time
Temperature
Pressure
DXPclave 18
DXPclave 24
B UNIVERSAL 4’ 3 10’ B 134 °C
2,03
bar
59’
Unwrapped solid and
hollow loads
5 kg 7 kg
Wrapped solid and
hollow loads
3,5 kg 5 kg
Double wrapped solid
and hollow instruments
2 kg 3,5 kg
Unwrapped porous
instruments
1 kg 2 kg
Wrapped porous
instruments
1 kg 2 kg
Double wrapped
porous instruments
1 kg 2 kg
B PRION 18’30” 3 14’ B 134 °C
2,03
bar
78’
Unrapped solid and
hollow instruments
5 kg 7 kg
Wrapped solid and
hollow loads
3,5 kg 5 kg
Double wrapped solid
and hollow instruments
2 kg 3,5 kg
Unwrapped porous
instruments
1 kg 2 kg
Wrapped porous
instruments
1 kg 2 kg
Double wrapped
porous instruments
1 kg 2 kg