Alarms
6. Alarms
6.1 Generality
When the device is switched on and during each sterilization cycle, the characteristic parameters of the various
phases of the cycle are constantly monitored, together with the proper functioning and acceptable condition of all
the components.
Any anomaly or fault is immediately indicated on the display through warning messages and alarm codes, together
with an acoustic signal.
The following paragraphs show the possible information, notification and alarm messages in the table:
INFORMATION: the information will appear in case of incorrect use or in case of maintenance requested by the
operator. These messages normally allow the cycle to start but require certain actions by the operator.
NOTIFICATIONS: warnings will appear in case of small malfunctions of the device; this type of malfunction does
not cause the sterilization cycle to be interrupted. The cycle will continue until completion but at the end of the
cycle it will be necessary to evaluate / solve the problem.
ALARMS: the alarms will appear in case of machine malfunctions that affect the sterilization cycle and lead to
“Sterilization process failed” according to EN 13060:2014+A1:2018. These malfunctions require the immediate
interruption of the process and the procedure for resetting the device . No further cycles can be performed until
the cause of the fault is resolved..
The tables show:
in the first column from the left the encoding of the message;
in the second column the message shown on the display;
in the third column the cause of the message;
in the fourth column the solution to the problem, which in some cases can be solved by the operator, in others
technical assistance is required.
Instruments and sterilization chamber are very hot.
Carry out the operations indicated without letting the parts of the body come into contact with hot
surfaces.
6.2 Informations list
NR. DISPLAY MESSAGE CAUSE SOLUTION
i01
Replace the bacteriological
filter
The bacteriological filter has
performed more than 500
cycles.
Replace the bacteriological filter
and reset the dedicated counter.
i02 Replace door gasket
The door gasket has performed
more than 1000 cycles.
Replace the door gasket and
reset the dedicated counter.
i03 Carry out periodic inspection
A year has passed since
installation date or more than
1000 cycles have been completed
without any maintenance by
technical support.
Require a complete check-up by
a qualified technical service;
contact technical maintenance.
i10 Close the door Door not closed properly.
Check that the door is closed
properly.
i20 Fill the tank
Insufficient water level in the
clean water tank.
Fill the clean water tank.
i21 Clean water tank full
The level of the clean water
tank has reached the maximum
allowed level.
Stop filling the clean water tank.
i22 Used water tank full The used water tank is full. Empty the used water tank.