17
Characteristics
3.8 Technical specifications
DXPclave 18 DXPclave 24
Chamber dimensions
Diameter 265mm 265mm
Depth 382mm 475mm
Chamber volume 18 lt 24 lt
Maximum load
(including trays)
Solid load 5,0 kg 7,0 kg
Porous load 1,5 kg 2,0 kg
Heating time
6’30” from room temperature
1’30” from preheated chamber
Sterilization time
B Universal 4’ 4’
B Prion 18’30” 18’30”
B Fast 3’30” 3’30”
B121 20’30” 20’30”
S Fast 3’30” 3’30”
Drying time
B Universal 10’ 10’
B Prion 14’ 14’
B Fast 5’ 5’
B121 16’ 16’
S Fast 7’ 7’
External dimensions 451 (W) x 492 (H) x 669 (L)
Net weight 59 kg 61 kg
Power voltage 230 V~ ±10 %
Frequency 50/60 Hz
Maximum power consumption 2350 W (10,7 A)
Average consumption 295 W (1,35 A)
Standby consumption 50 W (0,2 A)
Fuses 2 x T 16A, 400V (6,3x32mm)
Clock battery Internal, not replaceable by the operator: CR2032
Double water tank
5 l (clean water tank)
5 l (used water tank)
“Average” water consumption for standard cycles
134°C - 121°C - 3 vacuum
600 cc ÷ 700 cc 800 cc ÷ 1050 cc
Vacuum pump 13 l/min - 0.96 bar
Bacteriological filter 0.3 µm al 99.97 %
IP rating (in accordance with EN 60529) IP20 [1]
Separate heating system Band heater with separate power distribution.
Heat transmitted to the environment at 23°C 2,16MJ