18
Characteristics
Sound emission 60 dB(A)
Operating cycle Continuous
Pollution level 2 (IEC 61010-1)
Transient overvoltage II (IEC 61010-1)
Water conductivity control
LOW/HIGH WATER CONDUCIVITY (in reference to a
value of 15 microsiemens)
Available volume on trays 10 l
Maximum chamber temperature 137°C (-0+2°C)
Safety valve intervention pressure [2] 2,55 bar
[1] Classification of the device with respect to the ingress of liquids (second digit of the IP20 degree, or 0), and
protection from dangerous parts or against the access of solid foreign bodies (first digit of the IP20 degree, or 2).
The first characteristic digit indicates that:
2 - Degree that the casing provides for the protection of people against access to dangerous parts (finger of
test), preventing or limiting the penetration of a body part or a contested tool into the casing
by a person (test gauge with a diameter of 12 mm) and at the same time the degree that the casing provides
for the protection of the equipment against the penetration of solid foreign bodies (test sphere of diameter
12.5 mm);
The second characteristic figure indicates the degree of protection of the casing against harmful effects
on the equipment due to the penetration of liquids into the casing.
0 - no protection
[2] Pressurized container conforming to Directive 2014/68/UE (PED)