38
3.2 Alarm indicator
OER-Elite OPERATION MANUAL
Ch.3
Disinfectant replacement indicator
The indicator lights when the following reaches set value.
• The number of days of disinfectant usage
• The number of cycle count of disinfectant usage
• The number of day of shelf-life reaches
For how to set the disinfectant solution counter, refer to Section 4.12, “LCG replacement
indicator”.
For how to set the shelf-life, refer to Section 4.13, “LCG lot number and shelf-life management”.
Be sure to check the concentration of the disinfectant solution by using the
separately available test strips. The disinfectant replacement indicator does not
determine the effectiveness of the disinfectant solution. The expiration of the
disinfectant solution varies depending on many factors including the time since
activation, the temperature of the environment where the reprocessor is installed,
and the number and types of endoscopes that have been reprocessed. The
disinfectant replacement indicator does not take these and other factors into
consideration.
Detergent replacement indicator
The indicator lights when the following reaches set value.
• The number of days of detergent usage
• The number of day of shelf-life reaches
For details on the setting change, refer to Section 4.10, “Detergent replacement indicator”.
For how to set the shelf-life, refer to Section 4.11, “Detergent lot number and shelf-life
management”.
Filter replacement indicator
The lamp light when the number of days of filter usage reaches set value.
The number of days of water filter usage.
The number of days of air filter usage.
The number of days of gas filter (lid) usage.
The number of days of gas filter (tank) usage.
For details on the setting change, refer to Section 4.14, “Filter replacement indicator”.