Chapter 5 Operation
95
HIGH FLOW INSUFFLATION UNIT UHI-4
(5) Disconnect the suction tube from the CO
2
suction control pinch valve
of the high-flow insufflation unit (see Figure 5.9).
Figure 5.9
While pushing down the pinch valve with a finger, disconnect
the suction tube by an action like twisting it.
3. Press the insufflation switch of this instrument to perform insufflations for
exhausting the CO
2
from inside the instrument. When the CO
2
in the
instrument runs out, the insufficient gas supply caution is generated and the
insufflations stops automatically.
4. When this instrument enters stop mode, switch it OFF.
The illumination around the power switch turns off.
The front panel display also turns off.
5. Unplug the power cord from the hospital-grade power outlet.
6. Care for and store the high flow insufflation unit by referring to Chapter 6,
“Reprocessing, Storage, and Disposal”.
Pinch valve
Suction tube