Chapter 5 Operation
101
HIGH FLOW INSUFFLATION UNIT UHI-4
1. While holding the cavity mode selector switch, switch this instrument ON to
start it in the user mode (see Figure 5.10).
All the LEDs should light up when the power switch is set to on. After
confirming that all the LEDs are on, press and hold the cavity mode selector
switch for two seconds. During the User Mode, the Volume indicator
displays “888” and blinks.
Figure 5.10
2. The currently selected menu number is displayed on the set pressure
indicator, and the set value corresponding to the menu number on the flow
rate indicator (see Figure 5.11).
3. Each press of the cavity mode selection switch can change the menu
number as follows:
“1” “2” “3” “4” 5” “1”
4. To change the set value, press the cavity pressure control switches “+” or
“–”.
5. To end the user mode, turn the Power switch OFF. The device can retain the
settings.
1. Press and hold the cavity mode selector switch.
2. Press the power switch.