Chapter 6 Reprocessing, Storage, and Disposal
HIGH FLOW INSUFFLATION UNIT UHI-4
ISO 15883-2 regulate that disinfection temperature give an
A0 of at least 600. OLY recommends that WD provide an A0
value of 3,000.
Do not use a program that exceeds 55°C in the cleaning
process and 93C in the disinfection process.
Do not use a program that exceeds 10 minutes in the
disinfection process.
Confirm that there are no sudden temperature fluctuations in
the program you have selected.
Olympus performed validation using a WD compliant with
ISO 15883-1, 2 requirements.
Step Description
1 Pre-cleaning (25C)
2 Cleaning (with detergent, 55C, 10 min)
3 Rinsing (with or without neutralizer)
4 Heating (hot water)
5
Thermal disinfection which serves as the final rinsing
(without detergent and neutralizer, 93C, 5 min)
6 Drying (95C, 10 min)
Table 6.5 WD program parameters
*1 Olympus tested the compatibility of accessories with the WD compatible with
ISO 15883-1, 2. The drying temperature and the drying duration were 110°C
and 20 minutes, respectively
*2
.
*2 This value may not be the actual temperature of the accessories being
processed in the chamber, but it is a setting value for the WD. The drying
temperature for which Olympus confirmed material compatibility is 110°C as
the WD setting value. In the drying process, the temperature around the
product can be affected by various factors (e.g., chamber size, loading
amount, layout of instruments in the chamber).