Chapter 6 Reprocessing, Storage, and Disposal
115
HIGH FLOW INSUFFLATION UNIT UHI-4
Detergent solution for Washer-Disinfector
Use a low-foaming, medical-grade, alkaline enzymatic detergent (20 –
55°C, <pH 10.8).
Use a detergent whose compatibility for cleaning accessories is certified
by the WD’s manufacturer, and which are approved by a competent
authority. Follow the instructions provided by the detergent
manufacturer.
The detergent shown in Table 6.6 was used for validation.
Neutralizer
If rinsing with neutralizers after cleaning/disinfection is
necessary, make sure to remove all traces of neutralizers by
using the water as described in the instruction manual for
WD.
Use neutralizers recommended in the instruction manuals for the WD
and detergent.
Follow the instructions provided by the neutralizer manufacturer
regarding concentration, temperature, contact time and expiration date.
Trade name Type Manufacture
neodisher
®
Mediclean forte
Alkaline enzymatic Dr. Weigert
Table 6.6 Detergent used for validation