132
Chapter 6 Reprocessing, Storage, and Disposal
HIGH FLOW INSUFFLATION UNIT UHI-4
Sterilizing Insufflation Tube (MAJ-590) and Suction Tube
(MAJ-591)
This section describes the methods sterilizing those accessories which are listed
in Table 6.1 as being compatible with steam sterilization (autoclaving).
Steam sterilization (autoclaving)
After cleaning as described in “Cleaning Insufflation Tube (MAJ-590) and
Suction Tube (MAJ-591)” on page 127, steam sterilize (autoclave) the tube
according to the procedures described below.
Inspect each equipment package for openings, tears or other
damage. If the equipment package is open or damaged, seal
the equipment in new a package and re-sterilize it as
described below.
Allow the packages to dry within the autoclave, using the
autoclave’s drying cycle (if available) or by opening the door
of the autoclave and allowing the packages to air-dry.
Handling a wet package can compromise its sterility.
Do not pinch the tubes and connectors when enclosing them
in the package.
The expected sterilization effect cannot be achieved if the
sterilization pack is torn or peeled off. In this case, replace
the pack with a brand-new pack and use it for the
sterilization.
The sterilization effect cannot be achieved if too many
devices are packed in the sterilization pack. Be sure to leave
sufficient room when packing the sterilization pack.
Thoroughly dry the Insufflation Tube and Suction Tube before
sterilization.
The results of sterilization depend on various factors such as
how the sterilized instrument was packed or the positioning,
method of placing and loading of the instrument in the
sterilization device. Please verify the sterilization effects by
using biological or chemical indicators. Also follow the
guidelines for sterilization issued by medical administrative
authorities, public organizations or the infection management
sections at each medical facility, as well as the instruction
manual for the sterilization device.
After steam sterilization (autoclaving), let all components
gradually cool down to room temperature.