Appendix
147
HIGH FLOW INSUFFLATION UNIT UHI-4
Size Dimensions 375 (W) × 132 (H) × 380 (D) mm
(including projection parts)
370 (W) × 120 (H) × 330 (D) mm
(without projection parts)
Weight 9.3 kg
Classification
(medical
electrical
equipment)
Type of
protection
against
electric shock
Class I
Degree of
protection
against
electric shock
of applied part
TYPE CF applied part
Explosion
proofing
degree
Use in a flammable atmosphere is prohibited.
Maximum supply pressure
rating
16.2 MPa
Applicable gas CO
2
for medical use.
Connection to gas cylinder via the Olympus
cylinder hose (optional; five hose types available
to fit the specific CO
2
cylinders used: MAJ-1080
(PIN), MAJ-1081 (DIN), MAJ-1082 (ISO),
MAJ-1985(PIN), MAJ-1986(DIN)
Connection to medical gas pipeline via the
Olympus medical gas pipeline adapter
(MAJ-1084 (NIST), MAJ-1085 (DISS))
Pressure range of medical gas pipeline
–upper limit: according to ISO 7396 (1400 kPa)
–lower limit: 343.2 kPa (3.5 kgf/cm
2
)
Hose for medical gas pipeline is required to
comply with ISO 5359 (NIST or DISS)
Item Specification