148
Appendix
HIGH FLOW INSUFFLATION UNIT UHI-4
Supply pressure indications Five steps by LEDs.
5: above 4.5 MPa
4: 3.5 – 4.4 MPa
3: 2.5 – 3.4 MPa
2: 1.5 – 2.4 MPa
1: 0.1 – 1.4 MPa
Alarm:
When the cylinder pressure is lower than
2.5 MPa, an intermittent guidance sound
(1 second) will sound.
When the cylinder pressure is lower than
1.5 MPa, an guidance sound will sound again.
When the cylinder pressure is below 0.1 MPa,
CO
2
insufflation will stop, the cylinder pressure
indicator will light yellow and an intermittent alarm
(3 second) will sound.
Alarm
functions
Alarm list Excessive pressure Medium priority
Tube obstruction Low priority
CO
2
pressure 0 Low priority
Sound volume Medium priority 67 10 dB
Low priority 65 10 dB
Frequency/
duration
Medium priority 666 Hz for 1 s
Low priority 666 Hz for 0.4 s
Start/stop Operation
mode (start
switch ON and
lamp for start
lights)
The following functions are available:
CO
2
insufflation
Automatic suction/relief/smoke evacuation
(optional)
Cylinder pressure, flow rate indicators
Stop mode
(stop switch
ON and lamp
for stop lights)
Above operations, CO
2
insufflation and automatic
suction/smoke evacuation stop.
Settings, cylinder pressure, and volume displayed.
Selection of the cavity mode Select the cavity mode.
NORMAL” cavity mode for endoscopic
observation and treatment of the abdominal
cavity.
SMALL” cavity mode for observation and
treatment of the small cavity
(Example: colon (rectum)).
Cavity pressure control When cavity pressure reaches the set pressure
level, insufflation will stop. When cavity pressure
drops below the set pressure level, insufflation will
start.
Item Specification