150
Appendix
HIGH FLOW INSUFFLATION UNIT UHI-4
Flow rate
indications
Setting
indications
Flow rate indicator: Deep green indication of the set
pressure.
Actual value
indication
Flow rate bar graph: Green indicates the CO
2
flow
(instantaneous flow rate).
(All LEDs up to instantaneous flow rate value will
light.)
Flow rate indicator: White indicates the flow rate
value.
Flow rate
settings
3 flow setting
modes:
(“NORMAL”
cavity mode)
“HIGH” mode: 20 – 45 L/min
“MED” mode: 1.1 – 1.9 L/min, 2 – 19 L/min
“LOW” mode: 0.1 – 1.0 L/min
3 flow setting
modes:
(“SMALL”
cavity mode)
“HIGH” mode: 6 – 10 L/min
“MED” mode: 1.1 – 1.9 L/min, 2 – 5 L/min
“LOW” mode: 0.1 – 1.0 L/min
Previous flow
rate setting is
retained.
When power is turned ON, “LOW” mode is
selected.
VOLUME
indicator
Display
method
• 0.0 – 19.9 L, 0.1 L step
• 20 – 999 L, 1 L step
Reset Pressing the reset button will return the display to
0.0 L.
Smoke evacuation function
(optional)
• When the foot switch (optional) is depressed,
insufflation and suction are interlocked to
maintain cavity pressure at the set pressure while
smoke evacuation is performed.
• Combining ESG-400, USG-400, SonoSurg-G2,
UES-40 allows the automatic smoke exhaustion
to perform.
• This function will not perform in the following
conditions:
1. when the insufflation mode is LOW
2. when the cavity pressure is lower than about
2 mmHg
3. when insufflation is stopped
4. when an excessive pressure alarm is sounding
5. when an tube obstruction alarm is sounding, or
6. when the smoke exhaustion mode is OFF
Selection of the smoke
evacuation mode
Select the smoke evacuation mode.
“HIGH” mode: flow rate 10 L/min
“LOW” mode: flow rate 5 L/min
“OFF” mode: function off
Item Specification