Chapter 3 Installation and Connection
57
HIGH FLOW INSUFFLATION UNIT UHI-4
• Do not bend, pull, or twist the power cord. Equipment
damage including separation of the power plug and
disconnection of the cord wire as well as fire or an electric
shock can result.
• Be sure to connect the power plug securely to prevent
erroneous unplugging during use. Otherwise, the equipment
will not function.
• Do not extend a single wall mains outlet into multiple outlets
for connecting the power cords of both the electrosurgical
unit and light source. Otherwise, malfunction of the
equipment may result.
Confirm that the power is OFF.
1. Confirm that the power switch on the left part of the front panel is not in the
pushed-in position.
2. Connect the power cord of the high flow insufflation unit to the AC power
socket on the rear panel of the high flow insufflation unit (see Figure 3.16).
Figure 3.16
Properly
grounded wall
mains outlet
Power plug