62
Chapter 3 Installation and Connection
HIGH FLOW INSUFFLATION UNIT UHI-4
Connecting a suction tube (MAJ-591)
1.
Place a sterilized differential diameter connector over the end of the
sterilized small diameter tube (transparent). Connect the end of the
sterilized tube with the green line to the other end of the differential diameter
connector (see Figure 3.23).
2. Attach the sterilized small luer lock connector to the patient side of the tube
(see Figure 3.23).
Figure 3.23
3. If required, use the Y-shaped connector/extension tube to branch the
suction (see Figure 3.24).
Figure 3.24
4. Connect the small luer lock connector of the sterilized suction tube to the
connector on the trocar that is positioned close to where the smoke is being
created.
To suction
container
Patient side
Differential diameter
connector
Small luer lock
connector
Small diameter tube
Green-line tube
Patient side
To suction
container
Differential diameter
connector
Small luer lock
connector
Y-shaped
connector
Small diameter tube
To other suction device
Green-line tube