ZX-E
Operation Manual
Section 3 Adjusting Linearity
Section 3
BASIC OPERATION
Entering Adjustment Value for 50% Position
4. Move the sensing object to the 50% position.
5. Use the LEFT and RIGHT Keys to display
D050.
6. Press the ENT Key.
The sub-display will show OK and the adjustment value
will be registered.
Press the ENT Key again to reinput the adjust-
ment value.
Entering Adjustment Value for 100% Position
7. Move the sensing object to the 100% posi-
tion.
8. Use the LEFT or RIGHT Keys to display
D100.
9. Press the ENT Key.
The sub-display will show OK and the adjustment value
will be registered.
Press the ENT Key again to reinput the adjust-
ment value.
Rated measure-
ment distance
50%
Rated measure-
ment distance
100%
Z166-E1-02A.book Page 57 Friday, June 3, 2005 4:18 PM