Section 5 Changing Display Scales
ZX-E
Operation Manual
Section 5
DETAILED SETTINGS
Changing Display Scales
Change the display scale when you want to display a value different on the main display to the
actual measured value.
Place the sensing object in position and make the settings for either one-point or two-point
scaling.
The scaling set here is reflected in the display only. The output does not change.
The minimum display value is
19,999 and the maximum is 59,999. If the measurement result is less than
the minimum display value after scaling is executed, the display will be
19,999. If the measurement result is
greater than the maximum, the display value will be 59,999.
Scaling Application Example
Purpose Settings Method
To display distance from Sensor Head Offset the display value. One-point scaling
with no inverted
display values
p. 92
To display the height of the sensing object Offset the display value and invert the relation-
ship between display and reference values.
One-point scaling
with inverted dis-
play values
p. 92
Value to display:
2.5 mm
Sensor
Head
Sensing
object
Actual
measured
value:
1.0 mm
Display value (mm)
Actual measured
value (mm)
Value to display:
1.5 mm
Sensor
Head
Sensing
object
Actual measured value:
1.0 mm
Display value (mm)
Actual measured
value (mm)