Section 3 Adjusting Linearity
ZX-E
Operation Manual
Section 3
BASIC OPERATION
Entering Adjustment Values
Set the sensing object at the positions for 0%, 50%, and 100% of the rated measure-
ment distance and register those measured values as the adjustment values.
Measurement distance, p. 143
Perform the registration in order, starting with 0%, then 50% and 100% of the measure-
ment distance.
To simply replace the sensing object material and use the default linearity adjustment, skip this oper-
ation and perform Executing Adjustment.
Entering Adjustment Value for 0% Position
1. Set the sensing object at the 0% position.
2. Use the LEFT and RIGHT Keys to display
D000.
3. Press the ENT Key.
The sub-display will show OK and the adjustment value
will be registered.
Press the ENT Key again to reinput the adjust-
ment value.
Rated measurement distanc
e
0%
50%
100%
0%
Rated measure-
ment distance