80
Section 5 Setting Number of Samples to Average
ZX-E
Operation Manual
Section 5
DETAILED SETTINGS
Setting Number of Samples to Average
The number of samples to average is the number of data points used to average data
measured by the Sensor. The average value will be output.
Use the number of samples to average function to ignore sudden variations in measured
values. If the number of samples is increased, however, the response time of the judgement
outputs and linear output will be increased.
Moving to FUN mode and AVE
1. Set the mode switch to FUN.
2. Use the LEFT and RIGHT Keys to display
AVE on the main display.
Selecting Number of Samples to Average
3. Press the UP or DOWN Key.
The sub-display will flash.
4. Use the UP and DOWN Keys to select the
number of samples to average.
5. Press the ENT Key to confirm the selection.
The setting will be registered.
Selection for No. of
samples to average
Response time
1 0.3 ms
2 0.5 ms
4 0.8 ms
8 1.5 ms
16 2.5 ms
32 5 ms
64 (default) 10 ms
128 20 ms
256 40 ms
512 75 ms
1024 150 ms
2048 300 ms
4096 600 ms
Z166-E1-02A.book Page 80 Friday, June 3, 2005 4:18 PM